Product Mycotoxins Rapid Test kit

LSY-20007 Total Aflatoxins rapid test card 50T

Total Aflatoxins rapid test card

Catalog No. LSY-20007

1. Brief

This product is used for testing Total Aflatoxins(including B1, B2, G1,G2 etc.) in grain, feed, peanut and edible oil qualitatively, the whole test procedure only need 10-15min, easy to operate with high sensitivity.


2. Detection limit (ng/g or ppb)

Sample

Detection limit

Sample

Detection limit

Rice

3-5ppb

Peanut

3-5ppb

Wheat

3-5ppb

Soybean

3-5ppb

Corn

3-5ppb

Wheat bran

20-30ppb

Regular feed

5-8ppb

Rice bran

20-30ppb

DDGS

20-30ppb

Green fodder

20-30ppb

Soybean meal

10-15ppb

Edible oil

5-20ppb











3. Specifications:50 tests per box

4. Principle

The Total Aflatoxins rapid test strip is based on competitive inhibition immuno-chromatographic principle. In the flow process, Aflatoxins in the sample combined with Aflatoxins specific colloidal gold-labeled monoclonal antibody, Inhibit the combination between antibody and TAF -BSA conjugate on Test line of NC membrane, lead to the color change of Test line.

When the sample has no Aflatoxins residue or concentration lower than detection limit, T line is darker than or has same color with C line. when the concentration is equal to or higher than detection limit, T line is lighter than C line obviously or T line is invisible. No matter whether there is Aflatoxins residue in sample, C line will appear, it means the test is valid.  

5. Contents

Total Aflatoxins test card

50 pieces

Micro-well reagent

50 pieces

Manual

1 piece

TAF Sample extracting buffer

500ml







6. Sample preparation

6.1  A. Grain: rice, wheat, corn, sorghum

     B. Feed: regular feed, DDGS, soybean meal

     C. Beans: peanut, soybean

Weigh crushed sample 1±0.05 g into 15ml centrifuge tube, add 4ml TAF Sample extracting buffer, shake vigorously by hand or Vortex for 1-2 min, centrifuge at 4000r/min at room temperature for 5min, Take up-layer clear liquid to test.

6.2  Special sample ( sample is highly absorbent)

     Wheat bran, rice bran, green fodder

Weigh crushed sample 1±0.05 g into 15ml centrifuge tube, add 10ml TAF Sample extracting buffer, shake vigorously by hand or Vortex for 1-2 min, centrifuge at 4000r/min at room temperature for 10min, Take up-layer clear liquid to test.

6.3 Edible oil

Use Pipette to take 200ul sample into 50ml or 15ml centrifuge tube, add TAF Sample extracting buffer(Quantity as following tablet), shake vigorously by hand or Vortex for 3 min,take sample to test.

Detection limit (ppb)

5

10

15

20

B1 Sample extracting buffer (ml)

2ml

4ml

6ml

8ml

7. Operation procedures

7.1 Read the instruction carefully before use. Return test strips and sample into room temperature.

7.2 Take bottles needed from the kit package, take out required micro-wells and strips, making proper marks. Please use these test strips within 1h. Seal the cap of the bottles, avoid moisture.

7.3 Take 200ul of the test samples into the micro-wells, then repeatedly absorb for 5 times, mix the sample with the reagent in the micro-wells completely until no solid judge by eyes (this is a very important step), Incubate for 5min at room temperature(20-25℃).

7.4 Put test card on desk flatly, drop 80µl liquid from the micro-well, add into the “S” hole on test card.

7.5 Wait for 5-8min then read the result, it is invalid in other time.

8. Test Result Interpretation

8.1 Negative: T line is darker than or has same color with C line. It means there is no Aflatoxins residue in sample or the residue is lower than detection limit.

8.2 Positive: T line is obviouly lighter than T line or T line is invisible. It means the Aflatoxins residue is higher than or equal to detection limit.

9. Precautions

1) The test strips can be used only once at room temperature, do not use test strips out of expiry date.

2) Please seal the bottle after taking out required test strips. If can not use all 8 micro-wells at once, cup the rest and put it back to bottle and sealed.

3) Do not touch the white membrane surface in the middle of test strips, avoid sunlight and fan blowing directly.

4) Use the test strips testing again for positive results.

5) Please contact the supplier for any questions.

6) The result is for reference only, for confirmation, please refer to the relevant national standard method.

10. Specificity

Test 500ppb of Tetracyclines, Chloramphenicol, Sulfonamides etc., the result is Negative.

11. Storage and expiry date

Storage:  Store at 2-8 ℃ in dark, sealed, dry place, no frozen.

Expiry date: 12 months; date of production is on box.


Shenzhen Lvshiyuan Biotechnology Co., Ltd

D Building, National Biological Industrial Park of Marinelife, No.2 Binhai Road, Dapeng, Shenzhen, 518120 China

Tel. 86-755-28438788    

Fax 86-755-28938800

Email: info@lsybt.com     

Website: www.lsybt.com


Contact UsClose

Company: Shenzhen Lvshiyuan Biotechnology Co.,Ltd

WeChat/WhatsApp: +86-13427908554

Mobile: +86-18165709090 Skype: bellazou3

E-mail: info@lsybt.com, lsy@lsybt.com

Scan Codeclose