Product For Tissue(Fish, Shrimp, Chicken, Pork, Duck etc.)

LSY-20052 Nitrofuran(AHD) rapid test dipsticks (Tissue)

Nitrofuran(AHD) rapid test dipsticks (Tissue)

Catalog No. LSY-20052

1. Specifications

96 strips/kit

2. Brief

This product is used for testing Nitrofuran(AHD) in tissue sample qualitatively. Easy to operate, high sensitivity.

3. Sample performance

Poultry and livestock tissue: pork, beef, sheep, chicken, duck, goose, and other tissue;

Fish, shrimp and crab: Freshwater fish, marine fish, shrimp, crab and other tissue.

4. Detection limit

Tissue:  0.5μg/kg( ppb)

5. Contents

1

Nitrofuran(AHD) test dipsticks (include test dipsticks, desiccant, micro-wells reagent)

8 pieces/bottle, 12 bottles

2

Manual

1 piece

3

Reagent A (50mL/bottle) 1M HCl

3 bottles

4

Reagent B (30mL/bottle) Derivatization reagent    

1 bottle

5

Reagent C (100mL/bottle) 1M K2HPO4 solution

7 bottles

6

Reagent E (50mL/bottle) n-Hexane

2 bottles

7

Reagent F (100mL/bottle) Sample Dilution buffer

1 bottle

6. Principle

The Nitrofuran(AHD) rapid test dipstick is based on competitive inhibition immuno-chromatographic principle. In the flow process, Nitrofuran(AHD) in the sample combined with Nitrofuran(AHD) specific colloidal gold-labeled monoclonal antibody, Inhibit the combination between antibody and Nitrofuran(AHD)-BSA conjugate on Test line of NC membrane, lead to the color change of Test line. When the sample has no Nitrofuran(AHD) residue or concentration lower than detection limit, T line is darker than C line or T line has same color with C line; when the concentration is equal to or higher than detection limit, T line is obviously lighter than C line or T line has no color. No matter whether there is Nitrofuran(AHD) residue in sample, C line will appear, it means the test is valid.

7. Sample preparation

1) Take 5±0.05g homogenized tissue sample into 50ml centrifuge tube, add 4.5ml deionized water、1.5ml  Reagent A and 300ul Reagent B, shake and mix for 2min, put at 75℃ water bath for 5min;

2) Add 7ml Reagent C, shake and mix for 1min, add 6ml ethyl acetate, shake and mix for 1min, centrifuge at 4000r/min for 5min;

3) Take 3ml up-layer liquid into a 10ml centrifuge tube, blow to dry at 75℃ by Nitrogen or air.

4) Add 1ml Reagent E, shake to dissolve the dry residue (if do not dissolve the residue completely, it may lead to the result unstable), then add 1ml Reagent F, shake to even for 30s, be stable until layered, absorb out the up-layer(if appear emulsion, centrifuge at 4000r/min for 5min);

5) Absorb 200ul down-layer liquid (water layer), ready to test. (Avoid absorb the up-layer liquid).

Note: The treated sample can be used for the detection of Nitrofurans(AMOZ, AOZ, AHD, SEM) at the same time.

8. Operation procedures

8.1 Read the instruction carefully before use. Return test dipsticks and sample into room temperature.

8.2 Take bottles needed from the kit package, take out required microwells and dipsticks, making proper marks. Please use these test dipsticks within 1h. Seal the cap of the bottles, avoid moisture.

8.3 Take 200ul of the test samples into the microwells, then slowly aspirate for 5-10 times, mix the sample with the reagent in the microwells completely until no solid judged by eyes (this is a very important step).

8.4 Incubate for 5min at room temperature (20-25℃) , then insert the test dipsticks into the microwells with the "MAX" end fully dipped in to the mixture solution.

8.5 Incubate for 5-8min at room temperature again, read the result, it is invalid in other time.

9. Test Result Interpretation

9.1 Negative: T line is darker than C line, or T line has same color with C line. It means there is no Nitrofuran(AHD) residue in sample or the residue is lower than detection limit.

9.2 Positive: T line is obviously lighter than C line or T line is invisible. It means the Nitrofuran(AHD) residue is equal to or higher than detection limit.

9.3 Invalidation: C line isn’t seen wine red. It means the test card is out of efficacy, out of date or improper operation. Please run the test again using another package. If the invalid tests keep happening, please contact the supplier.   


10. Precautions

1) The test card can be used only once at room temperature, do not use test card out of expiry date.

2) Do not repeat use the tips to avoid cross pollution.

3) Do not touch the white membrane surface in the middle of test card, avoid sunlight and fan blowing directly.

4) Wear gloves when testing. 

5) Please contact the supplier for any questions.

11.Specificity

There is no cross reaction with Nitrofuran AMOZ, AOZ and SEM.

12. Storage and expiry date

Storage:  Store at 2-8 ℃ in dark, sealed, dry place, no frozen.

Expiry date: 12 months; date of production is on box.

Shenzhen Lvshiyuan Biotechnology Co., Ltd

D Building, National Biological Industrial Park of Marinelife, No.2 Binhai Road, Dapeng, Shenzhen, 518120 China

Tel. 86-755-28438788   Fax 86-755-28938800

Email: info@lsybt.com   Website: www.lsybt.com

Contact UsClose

Company: Shenzhen Lvshiyuan Biotechnology Co.,Ltd

WeChat/WhatsApp: +86-13427908554

Mobile: +86-18165709090 Skype: bellazou3

E-mail: info@lsybt.com, lsy@lsybt.com

Scan Codeclose