Malachite green rapid
test card (Water:2ppb)
Catalog No. LSY-20005-1W
1. Specifications
20 pieces/kit
2. Brief
This product is used for testing Malachite green
or Leucomalachite Green in water
sample (including pond water, etc.) qualitatively. Easy to operate, high sensitivity.
3. Detection limit
Water: 2ng/ml(ppb)
4. Contents
1
|
Malachite green
rapid test card (include test card, desiccant, micro-wells reagent)
|
20 T
|
2
|
Manual
|
1 piece
|
3
|
Reagent 1 (6ml/bottle)
|
1 bottle
|
5. Principle
The
Malachite green rapid test card is based on competitive inhibition immuno-chromatographic principle. In the flow process, Malachite green in the sample combined with Malachite green
specific colloidal gold-labeled monoclonal antibody, Inhibit the combination
between antibody and Malachite green-BSA
conjugate on Test line of NC membrane, lead to the color change of Test line.
When the sample has no Malachite green residue or concentration lower
than detection limit, T line is darker than C line or T line has same color
with C line; when the concentration is equal to or higher than detection limit,
T line is obviously lighter than C line or T line has no color. No matter
whether there is Malachite green residue in sample, C line will appear, it
means the test is valid.
6. Materials required but not provided
1) Equipment: Centrifuge, Micropipette,
balance (a reciprocal sensibility of 0.01 g);
2) Consumables: 15ml centrifuge tube,
disposable tips.
7. Sample preparation
Water sample:
No need for treatment (if there are too many impurities, centrifuge at 4000
rpm for 5 minutes, or let it stand), directly take 1.8ml of water sample and
add 0.2ml of reagent 1, mix well and wait for testing.
8. Operation
procedures
8.1 Read the instruction
carefully before use. Return test cards and sample into room temperature.
8.2 Take out test bag,
open and take out required micro-wells, making proper marks. Please use them
within 1h. Seal the test bag, avoid moisture.
8.3 Take 120µl of the
test samples into the micro-wells, then repeatedly pump and suck for many
times, mix the sample with the reagent in the micro-wells completely until no solid judged by eyes (this is a very important step). Incubate
for 2min at room temperature (20-25℃).
8.4 Put the test card
flatly, absorb the 80µl liquid from the
micro-well, add into the “S” hole on test card.
8.5 Be static, read the
result in 5~8min, it is invalid in other time.
9. Test Result Interpretation
9.1 Negative: T line is
darker than C line, or T line has same color with C line. It means there is no Malachite
green residue in sample or the residue is lower than detection limit.
9.2 Positive: T line is obviously lighter than C line or T line is
invisible or T line appear green
or blue color.
It means the Malachite
green residue is equal to or higher than detection limit.
10. Precautions
1) The test card can be
used only once at room temperature, do not use test card out of expiry date.
2) Do not touch the
white membrane surface in the middle of test card, avoid sunlight and fan
blowing directly.
3) Test again for
positive result;
4) Please contact the
supplier for any questions.
11.Specificity
Test 500ppb
Tetracyclines, Quinolones and Sulfonamides etc. The result is Negative.
12. Storage
and expiry date
Storage: Store at 4-30 ℃ in dark,
sealed, dry place, no frozen.
Expiry
date: 12
months; date of production is on box.
Shenzhen Lvshiyuan Biotechnology Co., Ltd
D Building,
National Biological Industrial Park of Marinelife, No.2 Binhai Road, Dapeng,
Shenzhen, 518120 China
Tel.
86-755-28438788
Fax
86-755-28938800
Email: info@lsybt.com
Website: www.lsybt.com