Nitrofuran(SEM) rapid test dipsticks (Tissue)
Catalog No. LSY-20053
1. Specifications
96 strips/kit
2. Brief
This product is used for testing Nitrofuran(SEM) in tissue sample qualitatively. Easy to operate, high sensitivity.
3. Sample performance
Poultry and livestock tissue: pork, beef, sheep,
chicken, duck, goose, and other tissue;
Fish, shrimp and crab: Freshwater fish, marine fish, shrimp, crab
and other tissue.
4. Detection limit
Tissue: 0.5μg/kg( ppb)
5. Contents
1
|
Nitrofuran(SEM) test
dipsticks (include test dipsticks, desiccant, micro-wells reagent)
|
8 pieces/bottle, 12 bottles
|
2
|
Manual
|
1 piece
|
3
|
Reagent A (50mL/bottle) 1M HCl
|
3 bottles
|
4
|
Reagent B (30mL/bottle) Derivatization reagent
|
1 bottle
|
5
|
Reagent C (100mL/bottle)1M K2HPO4 solution
|
7 bottles
|
6
|
Reagent E (50mL/bottle) n-Hexane
|
2 bottles
|
7
|
Reagent F (100mL/bottle) Sample Dilution buffer
|
1 bottle
|
6. Principle
The
Nitrofuran(SEM) rapid test dipstick is based on competitive inhibition immuno-chromatographic principle. In the flow process, Nitrofuran(SEM) in the sample combined with Nitrofuran(SEM)
specific colloidal gold-labeled monoclonal antibody, Inhibit the combination
between antibody and Nitrofuran(SEM)-BSA
conjugate on Test line of NC membrane, lead to the color change of Test line.
When the sample has no Nitrofuran(SEM) residue or concentration lower
than detection limit, T line is darker than C line or T line has same color
with C line; when the concentration is equal to or higher than detection limit,
T line is obviously lighter than C line or T line has no color. No matter
whether there is Nitrofuran(SEM) residue in sample, C line will appear, it
means the test is valid.
7. Sample preparation
1) Take 5±0.05g homogenized tissue sample into 50ml
centrifuge tube, add 4.5ml deionized water、1.5ml Reagent A and 300ul Reagent
B, shake and mix for 2min, put at 75℃ water bath for 5min;
2) Add 7ml Reagent C,
shake and mix for 1min, add 6ml ethyl acetate, shake and mix for 1min,
centrifuge at 4000r/min for 5min;
3) Take 3ml
up-layer liquid into a 10ml centrifuge tube, blow to dry at 75℃ by Nitrogen or air.
4) Add 1ml Reagent E, shake to dissolve the dry residue (if do not dissolve
the residue completely, it may lead to the result unstable), then add 1ml Reagent F, shake to even for 30s, be stable until layered,
absorb out the up-layer(if appear emulsion, centrifuge at 4000r/min for 5min);
5) Absorb 200ul down-layer
liquid (water layer), ready to test. (Avoid absorb the up-layer liquid).
Note: The treated sample can be used for the detection of Nitrofurans(AMOZ,
AOZ, AHD, SEM) at the same time.
8. Operation
procedures
8.1 Read the instruction
carefully before use. Return test dipsticks and sample into room temperature.
8.2 Take bottles needed
from the kit package, take out required microwells and dipsticks, making proper
marks. Please use these test dipsticks within 1h. Seal the cap of the bottles,
avoid moisture.
8.3 Take 200ul of
the test samples into the microwells, then slowly aspirate for 5-10 times, mix
the sample with the reagent in the microwells completely until no solid judged
by eyes (this is a very important step).
8.4 Incubate for 5min at
room temperature (20-25℃) , then insert the test dipsticks into the microwells with the
"MAX" end fully dipped in to the mixture solution.
8.5 Incubate for 5-8min
at room temperature again, read the result, it is invalid in other time.
9. Test Result Interpretation
9.1 Negative: T line is
darker than C line, or T line has same color with C line. It means there is no Nitrofuran(SEM) residue in sample or
the residue is lower than detection limit.
9.2 Positive: T line is obviously lighter than C line or T line is
invisible. It means the Nitrofuran(SEM) residue is equal to or
higher than detection limit.
9.3 Invalidation: C line isn’t seen wine
red.
It means the test card
is out of efficacy, out of date or improper operation. Please run the test
again using another package. If the invalid tests keep happening, please
contact the supplier.
10. Precautions
1) The test card can be
used only once at room temperature, do not use test card out of expiry date.
2) Do not repeat use the
tips to avoid cross pollution.
3) Do not touch the
white membrane surface in the middle of test card, avoid sunlight and fan
blowing directly.
4) Wear gloves when
testing.
5) Please contact the
supplier for any questions.
11.Specificity
There is no cross
reaction with Nitrofuran AMOZ, AHD and AOZ.
12. Storage
and expiry date
Storage: Store at 2-8 ℃ in dark,
sealed, dry place, no frozen.
Expiry
date: 12
months; date of production is on box.